57
Menu Display
Cal record, Sensor diagram, Statistics
Cal record
Data of last calibration, suitable for
documentation to ISO 9000 and GLP
(Date, time, calibration method, sensor zero and
slope, rel. humidity for calibration in air)
Sensor diagram
The measured values are continuously
monitored during the measurement
process. The sensor network diagram
provides at-a-glance information
about critical parameters.
If a tolerance limit has been exceeded,
the respective parameter is flashing.
Values in gray: Monitoring switched
off.
Statistics
Indication of sensor data for the
First Calibration and the last
3 calibrations.
(Date and time of first calibration, sensor zero point
and slope, temperature, pressure and response time)
Cal record
Return
23.0 °C
3.4 %O2
Last calibration
Cal mode
Zero
07/01/10 10:29
+0.010 nA
Slope
Impedance
–050.0 nA
998.3 kΩ
Relative humidity
0051 %
Statistics
Return
22.7 °C
3.4 %O2
Zero
1st Cal
Diff
+0.000nA 07/01/10 10:20
+0.010nA 07/11/10 12:34
Diff
Diff
–0.020nA 07/12/10 10:35
+0.900nA 07/18/10 10:42
Slope
Automatic - Water
Sensor diagram
Return
23.0 °C
3.4 %O2
2 - Zero point
1 - Slope
4 - Not in use
3 - Sensocheck
5 - Response time
7 - Sensor wear
6 - Cal timer
Critical range –
“inner circle”
Value out of
tolerance
The tolerance can be
modified as required!
“Outer circle”
Value within tolerance
Info
Sensor wear monitor
23.0 °C
Return
Sensor operating time
Autoclaving cycles
3.4 %O2
SIP cycles
CIP cycles
Sensor wear
635 d
1 of 2
0 of 3
1 of 5
Sensor wear monitor (ISM)
In addition to the current sensor wear,
the sensor operating time as well as
the number of executed autoclav-
ing, CIP, or SIP cycles can be seen at a
glance.
* Menu is only displayed for ISM modules
when a valid ISM sensor is connected.