17
Wiring
OXY 3400(X)-067
The signal from an external pressure transmitter can be fed in through the
external current input (18, 19). This enables automatic pressure correction for
oxygen measurement.
Sensor type
Standard
SE 7*6 ...
VP6-ST cable
Traces 01
SE 7*7 ...
VP6-ST cable
Traces 0011)
ISM2)
2
cathode
Coax core
transparent
Coax core
transparent
Coax core
transparent
8
anode
Coax shield
Red
Coax shield
Red
Coax shield
Red
12
reference
n.c.
Blue
13
guard/RTD
Gray + Green
Gray + Green
Gray + Green
14
RTD
White
White
White
15
shield
Cable shield
Yellow/green
Cable shield
Yellow/green
Cable shield
Yellow/green
16
ISM data
K8S connection:
Coax core
(transparent)
17
ISM DGND
K8S connection:
Coax shield
(black)
18
- I-Input
19
+ I-Input
1) With Protos 3400(X) only
2) With Protos II 4400(X) from FRONT firmware version 02.xx.xx