Menu Display
Module diagnostics, Sensor monitor
Open the diagnostics menu
From the measuring mode:
Press
menu
key to select menu.
Select diagnostics using arrow keys,
confirm by pressing
enter
.
Then select Module OXY.
The Diagnostics menu gives an over-
view of all diagnostics functions avail-
able. Messages set as “Favorite” can
be called directly from the measuring
mode using a softkey.
To configure:
Parameter setting / System control /
Function control matrix.
Module diagnostics
Function test of internal components:
- Internal device communication
- Check of firmware (module)
- Factory settings, measured value
processing
Sensor monitor
Shows the current directly measured
by the sensor, the barometric pressure,
and temperature.
Important function for diagnostics and
validation!
Menu selection
Return to meas
23.0 °C
Select:
[enter]
Lingua
3.4 %O2
Module OXY 3400-067
Return
23.0 °C
3.4 %O2
Sensor monitor
Cal record
Sensor network diagram OXY
Set favorite
Module diagnostics
Module diagnostics
Return
23.0 °C
3.4 %O2
Communication
Flash check sum
Meas. processing
ok
ok
ok
Sensor monitor
Return
23.0 °C
3.4 %O2
Sensor current
Sensor current (25°C)
Air pressure
–40.03 nA
–40.03 nA
1014 mbar
Ext. pressure transmitter
RTD
0 bars
22.1 kΩ
Diagnostic Functions