37
Message
Sensocheck can generate a message
for failure or maintenance request.
It can be seen in the Message list of
the Diagnostics menu.
Setting the Sensor Data Parameters
Sensor data: Sensor monitoring details
Note:
Function check (HOLD) mode active
Menu
Module OXY 3400-067 (Administrator)
20.1 °C
Return
Pressure correction
Sensor data
Cal preset values
100.4%Air
Salinity correction
Messages
Input filter
Sensor data (Administrator)
20.1 °C
Abort
Sensoface
100.4%Air
Temperature probe
Display
Display
Parameter selection
Sensor type
Block
OK
Sensocheck (Administrator)
20.1 °C
Abort
100.4%Air
Nominal
(NTC 22kohm)
Off
On
Off
Failure
Maint. request
Message
Sensor monitoring details (administrator)
20.1 °C
Response time
Zero
Sensocheck
100.4%Air
Sensor wear
CIP counter
Slope
Return
(Auto)
(Auto)
(Auto)
(Off)
(Auto)
(Auto)
Min
Max
Monitoring
0100 kohms
0029 kohms
03
Auto
Sensoface
provides information on
the sensor condition. Great devia-
tions are signaled. Sensoface can be
switched off.
Sensor monitoring details
The following parameters are
monitored: slope, zero, response
time, temperature, for ISM sensors
also sensor wear, CIP/SIP counter,
autoclaving counter and sensor
operating time, membrane and inner
body changes. For “Auto”, the toler-
ance limits are displayed in gray.
For “Individual”, the settings can be
specified by the user.
Sensor data
(see following page)
Sensor data are preset depending on
the sensor type.
Gray display lines cannot be edited.
Polarization voltage
Sensor polarization
500/675 mV
Auto
ISM sensors automatically provide
most of the default settings.
Individual settings are not
overwritten.
Measure in
Gases
Liquids
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.