Protos OXY 3400(X)-067 Specifications
Input for sensors
SE 7*6 ... , SE 7*7 ... or "other"
Automatic range selection:
Input range 1
Meas. current 0 ... 600 nA, resolution 10 pA
Meas. error
2)
< 0.5 % meas. val. + 0.05 nA + 0.005 nA/K
Input range 2
Meas. current 0 ... 10000 nA, resolution 166 pA
Meas. error
2)
< 0.5 % meas. val. + 0.8 nA + 0.08 nA/K
Display ranges
Standard sensor
(SE 7*6 ...)
Trace sensor
01
(SE 7*7 ...)
Trace sensor
001
3)
Other
Saturation
(-10 ... 80 °C)
000.0 ... 999.9 %Air
0.000 ... 9.999 %Air
00.00 ... 99.99 %Air
000.0 ... 999.9 %Air
0.000 ... 9.999 %Air
00.00 ... 99.99 %Air
000.0 ... 999.9 %Air
0.000 ... 9.999 %Air
00.00 ... 99.99 %Air
000.0 ... 999.9 %Air
00.00 ... 99.99 %O2
000.0 ... 999.9 %O2
0.000 ... 9.999 %O2
00.00 ... 99.99 %O2
000.0 ... 999.9 %O2
0.000 ... 9.999 %O2
00.00 ... 99.99 %O2
000.0 ... 999.9 %O2
0.000 ... 9.999 %O2
00.00 ... 99.99 %O2
000.0 ... 999.9 %O2
Concentration
(-10 ... 80 °C)
(dissolved oxygen)
00.00 ... 99.99 mg/l
000.0 ... 999.9 mg/l
0000 ... 9999 µg/l
00.00 ... 99.99 mg/l
000.0 ... 999.9 mg/l
000.0 ... 9.999 µg/l
0000 ... 9999 µg/l
00.00 ... 99.99 mg/l
000.0 ... 999.9 mg/l
0000 ... 9999 µg/l
00.00 ... 99.99 mg/l
000.0 ... 999.9 mg/l
00.00 ... 99.99 ppm
000.0 ... 999.9 ppm
0000 ... 9999 ppb
00.00 ... 99.99 ppm
000.0 ... 999.9 ppm
000.0 ... 9.999 ppb
0000 ... 9999 ppb
00.00 ... 99.99 ppm
000.0 ... 999.9 ppm
0000 ... 9999 ppb
00.00 ... 99.99 ppm
000.0 ... 999.9 ppm
Volume concentration
in gas
00.00 ... 99.99 Vol%
000.0 ... 999.9 Vol%
0000 ... 9999 ppm
0.000 ... 9.999 Vol%
00.00 ... 99.99 Vol%
000.0 ... 999.9 ppm
0000 ... 9999 ppm
0.000 ... 9.999 Vol%
00.00 ... 99.99 Vol%
0000 ... 9999 ppm
0.000 ... 9.999 Vol%
00.00 ... 99.99 Vol%
000.0 ... 999.9 Vol%
Partial pressure
0000 ... 9999 mbar
00.00 ... 00.00 mbar
000.0 ... 000.0 mbar
0000 ... 9999 mbar
0.000 ... 9.999 mbar
00.00 ... 00.00 mbar
000.0 ... 000.0 mbar
0000 ... 9999 mbar
00.00 ... 00.00 mbar
000.0 ... 000.0 mbar
0000 ... 9999 mbar
0000 ... 9999 mmHg
00.00 ... 00.00 mmHg
000.0 ... 000.0 mmHg
0000 ... 9999 mmHg
0.000 ... 9.999 mmHg
00.00 ... 00.00 mmHg
000.0 ... 000.0 mmHg
0000 ... 9999 mmHg
00.00 ... 00.00 mmHg
000.0 ... 000.0 mmHg
0000 ... 9999 mmHg
Permissible guard current
≤ 20 µA
Polarization voltage
0 ... –1,000 mV, default –675 mV (resolution < 5 mV)
Pressure correction
Air pressure
700 ... 1100 mbar
Manual
0 ... 9999 mbar
External
0 ... 9999 mbar (through current input 0(4) ... 20 mA input)
Via bus
0 ... 9999 mbar (via PROFIBUS or Foundation Fieldbus)
Salinity correction
0.0 ... 45.0 g/kg