39
Parameter
Default
Selection / Range
Sensor monitoring details (continued)
Auto: Other
Nominal:
-0100 nA
(with ISM sensor: Default slope)
Min.:
-0900 nA
(with ISM: Min. range)
Max.:
-0030 nA
(with ISM: Max. range)
Individual:
permisible range 25 ... 9999 nA
Slope message
Maint. request
Off, failure, maintenance request
Zero
Auto
Auto, Individual
Auto: Standard sensor 01 (SE 7*6 ...)
Nominal:
0.000 nA
Min.:
-1.000 nA
Max.:
1.000 nA
Auto: Trace sensor 01 (SE 7*7 ...)
Nominal:
0.000 nA
Min.:
-1.000 nA
Max.:
+1.000 nA
Auto: Trace sensor 001
Nominal:
0.000 nA
Min.:
-3.000 nA
Max.:
+3.000 nA
Auto: Other
Nominal:
0.000 nA
Min.:
-1.000 nA
Max.:
1.000 nA
Individual:
permissible range -3.000 ... +3.000 nA