41
Parameter
Default
Selection / Range
Cal preset values
Product calibration
%Air
SAt (%Air), Conc (mg/l, µg/l, ppm, ppb),
p´ (mmHg, mbar)
Calibration timer
- Monitoring
Auto
Off, Auto, Individual
- Cal timer
0000 h
With ISM: Off, Without ISM: Entry (xxxx h)
Pressure correction
Pressure transmitter
Difference
Absolute, Difference
I input
4 ... 20 mA
0 ... 20 mA / 4 ... 20 mA
Start 0(4) mA
0000 mbar
xxxx mbar
End 20 mA
9999 mbar
xxxx mbar
Pressure during meas
Air pressure
Air pressure, Manual (default 1013 mbar), External
Pressure during cal
Air pressure
Air pressure, Manual (default 1013 mbar), External
Salinity correction
Input
Salinity
Salinity, Chlorinity, Conductivity
(00.00 g/kg or 0.000 µS/cm, depending on selection)
Messages (gas)
Concentration
Off
Off, Variable limits
Partial pressure
Off
Off, Variable limits
Air pressure
Off
Off, Device limits max., Variable limits
Messages (liquid)
%Air saturation message
Off
Off, Variable limits
%O2 saturation message
Off
Off, Variable limits
Concentration message
Off
Off, Variable limits
Partial pressure message
Off
Off, Variable limits
Air pressure message
Off
Off, Device limits max., Variable limits
Sensor devaluation (ISM)