23
Menu Display
Selecting a calibration method
Open calibration
Press
menu
key to select menu.
Select calibration using arrow keys,
press
enter
to confirm, passcode 1147
(To change passcode, select:
Parameter setting / System control /
Passcode entry).
Calibration:
Select “Module OXY”
Menu selection
Return to meas
83.0%Air
25.6 °C
Select:
[enter]
Lingua
Calibration
Return
25.6°C
Module PH 3400-032
Module CONDI 3400-051
Info
83.0%Air
Module OXY 3400-067
Module OXY 3400-067
Return
25.6°C
Automatic - Air
Data entry
Product calibration Sat
Zero correction
83.0%Air
Automatic - Water
Select a calibration method:
•
Automatic - Water
•
Automatic - Air
•
Product calibration Sat
(concentration/partial pressure)
• Data entry
•
Zero correction
•
Temp probe adjustment (with Protos II 4400(X))
When you activate calibration, the
analyzer automatically proposes the
previous calibration method.
If you do not want to calibrate,
”Return” with the left softkey.
During calibration the module is in
function check (HOLD) mode.
Current outputs and relay contacts
behave as configured (BASE module).