12
ISM
Predictive Maintenance
Module OXY 3400-067
Return
205 mbar
Adjust temp probe
Membrane body changes
Inner body changes
Autoclaving counter
80.7%Air
Sensor monitor
The settings for predictive maintenance are
made in the
•
Maintenance menu /
Module selection
Autoclaving counter
Max. cycles
Count cycles
205 mbar
80.7%Air
Return
1
050
007
Membrane body changes
Max. changes
Count changes
205 mbar
80.7%Air
Return
Take over
004
000
Update sensor data after change:
Serial number
09/06/10
Date
0000772
Inner body changes
Max. changes
Count changes
205 mbar
80.7%Air
Return
Take over
004
000
Update sensor data after change:
Serial number
09/06/10
Date
0000772
Autoclaving Counter
When setting the sensor data, the maximum
number of autoclaving procedures permit-
ted must be specified. Then, each cycle can be
recorded in the Maintenance menu. This shows
how many autoclaving cycles are still permitted.
Membrane Body Changes
During parameter setting, the maximum num-
ber of membrane body changes permitted must
be specified. Then, each cycle can be recorded
in the Maintenance menu (date, serial number).
This shows how many changes are still permit-
ted.
Inner Body Changes
During parameter setting, the maximum num-
ber of membrane body changes permitted must
be specified.
Then, each change of an inner body can be
recorded in the Maintenance menu (date, serial
number). This shows how many changes are still
permitted.
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.