47
Messages
Setting the Message Parameters
Messages
Note:
Function check (HOLD) mode active
Menu Display
Messages
All parameters determined by the
measuring module can generate
messages.
• Device limits max:
Messages are generated when the
process variable (e.g. air pressure) is
outside the measurement range.
The “Failure” icon is displayed, the
NAMUR failure contact is activated
(BASE module, factory setting:
contact K4, N/C contact). The current
outputs can signal a 22 mA message
(user defined).
• Variable limits:
For the “failure” and “warning”
messages you can define upper and
lower limits for message generation.
• Message icons:
Messages (Administrator)
20.1 °C
Return
83.4%Air
Messages Air pressure (Administrator)
20.1 °C
Abort
Monitoring
83.4%Air
Variable limits
OK
Device limits max.
Off
Messages Partial pressure
Messages Saturation %O2
Messages Saturation %Air
Message Air pressure
Messages Concentration
Messages (Administrator)
20.1 °C
83.4%Air
Warning Limit Hi
Monitoring
Failure Limit Lo
Failure Limit Hi
Warning Limit Lo
1070 mbar
Variable limits
1100 mbar
0730 mbar
0700 mbar
Abort
OK
Message list
22.3 °C
Fail O2 air pressure
83.4%Air
Fail O2 concentration
Abort
Diagnostics menu
When the “Maintenance” or “Failure”
icons are flashing in the display, you
should call up the Diagnostics menu.
The messages are displayed in the
“Message list”.
Failure (Failure limit HiHi/LoLo)
Maintenance (Warning limit Hi/Lo)