40
Parameter
Default
Selection / Range
Sensocheck
Auto
Auto, Individual
Monitoring of nominal impedance; determined by calibration, taken over by adjustment.
Individual
allows specifying the values:
Nominal:
xxxx kΩ
Min.:
xxxx kΩ
Max.:
xxxx kΩ
Sensocheck message
Maint. request
Off, failure, maintenance request
Response time
Auto
Auto, Individual
Auto:
max. 1200 sec
Individual:
xxxx sec
Response time message
Maint. request
Off, failure, maintenance request
Temperature
Auto
Auto, Individual
Auto:
Min.:
-020.0 °C
(with ISM: Min. measuring temp)
Max.:
+150.0 °C
(with ISM: Max. measuring temp)
Individual:
Min.:
xxx.x °C
Max.:
xxx.x °C
Temperature Message
Maint. request
Off, failure, maintenance request
Sensor monitoring details (ISM sensors only)
Sensor wear
Auto
Off, Auto, Individual
CIP counter
Off
Off, Individual (Enter max. CIP cycles)
SIP counter
Off
Off, Individual (Enter max. SIP cycles)
Autoclaving counter
Off
Off, Individual (Enter max. autoclaving cycles)
Sensor operating time
Off
Off, Individual (Enter max. operating time)
Membrane body changes
Off
Off, Individual (Enter max. changes)
Inner body changes
Off
Off, Individual (Enter max. changes)
N
ot
e: