AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 3
7
Rear label
7
Fig. 3:
Label
1.4
Performance features
The newly designed optical bench with advanced temperature control
ensures high sensitivity, fast baseline stabilization, and excellent repro-
ducibility. Furthermore, the long-life LED and highly back pressure resis-
tant flow cell guarantee minimal maintenance.
If the Extended Dynamic Range option is active, it is possible to broaden
the linear dynamic range.
The detector measures with a maximum data rate of 100 Hz. During the
measurement, approx. 100 spectra per second are recorded.
The detector can be operated with one of the available chromatography
data systems (OpenLAB® EZChrom Edition, ClarityChrom®, Chrome-
leon™, PurityChrom® and Mobile Control Chrom), as well with an option-
al touch display (Mobile Control), via LAN or analog input/output, by
which it can be integrated in nearly every LC system.
The Mobile Control and supported software products display or read
GLP data, such as the operating hours of the device or light sources. For
a detailed description of viewing or reading the GLP data, see the appro-
priate instructions for the software products:
https://www.knauer.net/en/Support/User-manuals/Software
).
Light intensity and balance, zero glass position and LED current can be
easily monitored via the Mobile Control and supported software products
to ensure optimal working conditions for your analysis.
Optical bench
Extended Dynamic Range
Data rate
Control
GLP data
Troubleshooting