AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 25
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
cau ses a short circuit.
Turn off the device before connecting cables.
Pull the power plug.
Electronic defect
Electrostatic discharge can destroy the electronics.
Wear a protective bracelet against electrostatic discharge and ground.
1.
Insert the operating tool
3
in an
upper small opening at the front
of the pin header
1
.
2.
Insert the cable into the open-
ing
2
underneath the inserted
operating tool.
3.
Pull out the operating tool.
Check whether the cables are
tightly attached. Press the pin
header onto the connector. Finish the installation. Bring the device into
operation.
6.6
Integrator port
The integrator output supplies the current signal value as an analog volt-
age (0–2.5 V). Via software or Mobile Control, the signal cal be scaled to
0,1 / 0,5 / 1 / 2 / 5 μRIU/mV (see 7.2.7 on page 30). The integrator out-
put is connected to other instruments with the analog connection cable
(Cinch-Cinch, supplied with delivery) or other special connection cables
(not supplied with delivery).
Note:
The multiple connnection to ground (simultaneous grounding via
GROUND and integrator port) must be avoided. It could induce noise
loops that may lead to disturbed measurement results.
For details on the integrator connector, see the list below:
non-bipolar
1 channel
0-2.5 V
DAC 20 bit
scalable
Adjustable to offset
2
3
1
Process
Next Steps
Connection specification