AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 19
The computer has been switched off.
There is a patch cable for each device and the computer.
1.
Use the patch cable to connect the router and the computer. Repeat
this step to connect all devices.
2.
Use the power supply to connect the router to the mains power
system.
6.4.3 Configuring the router
The router is preset at the factory. Information about IP address, user
name and password is mentioned in the instructions of the router:
https://goo.gl/ahGhmG
.
1.
To open the router configuration, start your Internet browser and enter
the IP address (not for all routers).
2.
Enter user name and password.
3.
Configure the router as DHCP server.
4.
In the router configuration, check the IP address range and make
changes if necessary.
Note:
If the IP address range should have been changed, make surte to
note his information on the router.
As soon as the router has assigned a n IP address to each device, the
chromatography software takes over the control of the chromatography
system.
6.4.4 Integrating the LAN into the company network
A network administrator can integrate the LAN into your company net-
work. In this case you use the WAN port of the router.
An additional patch cable is present.
1.
Check that the IP address range of the router and of the company
network do not overlap.
2.
In case of an overlap, change the IP address range of the router.
3.
Use the patch cable to connect the router WAN port to the company
network.
4.
Restart all devices, including the computer.
6.4.5 Controlling several systems separately in the LAN
Devices connected to a LAN communicate through ports, which are part
of the network address. If more than one chromatography systems are
connected to the same LAN and you plan on controlling them separate-
ly, you can use different ports to avoid interference. Therefore, the port
number for each device must be changed and this same number must be
entered into the device configuration of the chromatography software.
We recommend to use the same port number for all devices in the same
system.
Note:
The default port is set to 10001. You must use the same numbers
in the device configuration of the chromatography software as in the
device, otherwise the connection fails.
1.
Find out port number and change it on the device.
2.
Enter the port number in the chromatography software.
The connection is established.