AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
10
Product information
The following symbols and signs can be found on the device:
Symbols
Meaning
High-voltage hazard
Electrostatic discharge hazard, damages to sys-
tem, device, or components can occur.
Obey maximum load for leak tray during transpor-
tation, installation and operation.
A device marked with CE fulfills the product spe-
cific requirements of European directives. This is
confirmed in a Declaration of Conformity.
Testing seals in Canada and the USA at nationally
recognized testing centers (NRTL). The certified
device or system has successfully passed the qual-
ity and security tests.
Because of its general temperature sensitivity, the detector must always
be the lowest device in a system. The RI detector must always be situated
below a UV detector because it is more sensible to temperature.
The capillary kits for easy installation (available as accessory) only fit if the
system is set up accordingly.
Note:
At the time of delivery, the detector is filled with ethanol. Flush the
system (with the flush valve open) with water for 60 minutes before start-
ing the measurement.
5.1
Operating environment
Before you specify the location, read the technical data (see 11 on page
40). There you will find all the important information about power con-
nection, ambient conditions and humidity.
Only if the requirements for ambient conditions of the operating envi-
ronment are met, the intended use can be ensured. In order to ensure
thermo stability and to prevent drift effects, note the following aspects:
More effective aspects for detectors in HPLC systems are described in the
according chapter (see 6.3 on page 16).
Note:
The leak may malfunction it the device is placed on a tilted surface.
Check the horizontal position with a spirit level.
4. Symbols and signs
5. Unpacking and setup