AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
40
Product information
Procedure
Figure
1.
Using the open-end wrench,
disconnect the first capillary
and close the opening with a
hole plug.
2.
Using the open-end wrench,
disconnect the second cap-
illary and close the opening
with a hole plug.
3.
Disconnect the power cable
from the connector
1
on the
rear side of the detector.
If you want to ship the device, reuse the packaging materials that you
received with the delivery.
11.1 Main features
AZURA® Detector RID 2.1L
AZURA® Detector RID 2.1L HighFlow
Detector type
Analytical refractive index detector
Preparative refractive index detector
Light source
Long-life LED
Refractive index range
1.00–1.75 RIU
Noise
± 2.5 nRIU
± 50 nRIU
Drift
200 nRIU/h
2000 nRIU/h
Linearity
>1000 µRIU
>4000 µRIU
Temperature control
OFF, 30–55 °C (1 °C steps)
Max. flow rate
10 ml/min (pure water)
100 ml/min (pure water)
Time constants
0.00 / 0.01 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s
Autozero
Full range
Flow cell
5 bar backpressure resistance (flow cell included)
Flow cell volume
15 µl
9 µl
Wetted materials
Stainless steel/fused silica/PTFE/PEEK
GLP
Detailed report about lamp recognition, operating hours, lamp operating
hours, number of lamp ignitions.
Life-threatening injuries
Health danger if getting in contact with toxic, infectious or radio-active
substances.
Before disposing of the device or sending it away for repair, you are
required to decontaminate the device in a technically correct manner.