AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
Product information 11
Device defect
The device overheats at exposure to sunlight and insufficient air circula-
tion. Device failures are very likely.
Set up the device in such a way that it is protected against exposure
to direct sunlight.
Leave room for air circulation: See paragraph „space requirements“.
Set up the detector on an even surface.
Protect the detector from direct sunlight.
Set up the detector in a place protected from draft (air conditioning).
Protect the detector from strong draft.
Do not set up the device near to other machines that cause floor
vibrations.
Avoid vibration.
Keep the detector away from high frequency sources. High-frequency
sources may compromise measuring values.
If you are located in an earthquake area, use the bore holes
1
. The
bore holes are located on both right and left side panels.
Minimum of 5 cm with a device on one side.
Minimum of 10 cm with devices on both sides.
5.2
Unpacking the detector
The detector is delivered in a closed package. To prevent damage, the
package is the best possible protection for the protector.
You have checked the carton for shipping damage.
Utility knife
Bruising danger
Damage to the device by carrying or lifting it on protruding housing
parts. The device may fall and thus cause injuries.
Lift the device only centrally on the side of the housing.
General requirements
1
Space requirements
Prerequisites
Tools