17.3
Earthing the VESA Mount for Large Screens
•
Place the first contact washer D4,1 on
the cross recessed head screw
M4 x 16 mm DIN 7985.
•
Place the cross recessed head screw
through the drill hole in the VESA mount
for large screens
.
•
Place the second contact washer D4,1
on the cross recessed head screw .
•
Attach the earthing shoe of the earthing
cable and the third contact
washer D4,1 .
•
Attach and tighten a hexagonal nut M4
.
Make sure the earthing cable is safely in
place:
•
The earthing shoe of the earthing
cable must butt against the VESA
mount for large screens
.
•
The 2 contact washers D4,1 / must
be placed in front of and behind the
earthing shoe of the earthing cable .
•
The hexagonal nut M4 must be
tightened.
Fig. 17-3:
Earthing the VESA mount for large screens
•
Stick an earthing symbol (not illustrated) onto the VESA mount for large screens
next
to the earthing point.