6-4
Chapter 6. Calibration
The greater the known quantity of material
added during the calibration procedure, the
greater the accuracy of the calibration.
Adding 50% of the total capacity results in
greater accuracy than adding 25% of the total
capacity.
Follow this procedure to calibrate by adding a
known quantity of material.
1. If the Weigh II is in Auto Mode (Auto LED
illuminated), press the Auto/Man Key to
put the system in Manual Mode. The Auto
LED turns off.
2. Use the Up Arrow and Down Arrow Keys to
scroll to the desired vessel.
3. Press the Menu Key to display the
Main
Menu. The display shows:
4. Press the F3 Key to access the
Cal
Menu.
The display shows:
5. Press the F1 Key to access the
Auto
Menu. The display shows:
6. Press the F1 Key to access
LoSpan
. The
display looks like this:
The unit of measure you set up in the
Units
Menu shows in place of lbs.
7. Use the keypad or the Up and Down Arrow
Keys to input a value that represents the
estimated quantity of material in the vessel
as the
LoSpan
value. Press the Enter Key
to save the value in memory.
Lo Span
is
now set [point (1) in Figure 6-3]. The
display flashes a message acknowledging
your selection and returns to:
8. Add a known quantity of material, repre-
senting at least 25% of the vessel’s total
capacity, to the vessel.
9. Press the F3 Key to access
HiSpan
. The
display looks like this:
The unit of measure you set up in the
Units
Menu shows in place of lbs.
10.
Using the keypad or the Up and
Down Arrow Keys, input the value equal to
the sum of the known quantity (Step 8) and
the estimated quantity (Step 7). Press the
Enter Key to save the value in memory.
Hi
Span
is now set [point (2) in Figure 6-3].
The display flashes a message acknowl-
edging your entry and returns to:
11.
Press the Esc Key to scroll up the
menu tree or press the Auto/Man Key to
return the display to vessel monitoring.
Note
Shifting the line from the dashed line to
the solid line in Figure 6-3 is accom-
plished by setting zero. The procedure
to set zero is described in
Refining the
Calibration by Setting Zero
in this
section. Set zero when the vessel
contains a known quantity of material
(usually none) at a later date.
Calibration by Subtracting a
AUTO CAL MENU
LoSpan HiSpan
F1 F2 F3
HI SPAN AUTO CAL
> X lbs
F1 F2 F3
AUTO CAL MENU
LoSpan HiSpan
F1 F2 F3
WII MAIN MENU
Disp I/O Cal
F1 F2 F3
LO SPAN AUTO CAL
> X lbs
F1 F2 F3
AUTO CAL MENU
LoSpan HiSpan
F1 F2 F3
CALIBRATION MENU
Auto Manual
F1 F2 F3
Summary of Contents for Weigh II
Page 12: ...2 4 Chapter 2 Hardware Installation...
Page 20: ...3 8 Chapter 3 Menu Tree Keyboard Functions and Quick Start...
Page 50: ...6 12 Chapter 6 Calibration...
Page 68: ...A 2 Appendix A Product Specifications...
Page 70: ...B 2 Appendix B Summary of Commands Used When Vessel Mounting Hot Keys...
Page 77: ...Appendix E E 1 This appendix not used at this time...
Page 78: ...Appendix E E 2...
Page 80: ...Appendix F Technical Drawings F 2...
Page 81: ...Appendix F Technical Drawings F 3...
Page 82: ...Appendix F Technical Drawings F 4...
Page 83: ...Appendix F Technical Drawings F 5...
Page 84: ...Appendix F Technical Drawings F 6...
Page 85: ...Appendix F Technical Drawings F 7...
Page 86: ...Appendix F Technical Drawings F 8...
Page 87: ...Appendix F Technical Drawings F 9...
Page 88: ...Appendix F Technical Drawings F 10...
Page 89: ...Appendix F Technical Drawings F 11...
Page 90: ...Appendix F Technical Drawings F 12...
Page 91: ...Appendix F Technical Drawings F 13...
Page 92: ...Appendix F Technical Drawings F 14...
Page 93: ...Appendix F Technical Drawings F 15...
Page 94: ...Appendix F Technical Drawings F 16...
Page 95: ...Appendix F Technical Drawings F 17...
Page 96: ...Appendix F Technical Drawings F 18...
Page 97: ...Appendix F Technical Drawings F 19...
Page 98: ...Appendix F Technical Drawings F 20...
Page 99: ...Appendix F Technical Drawings F 21...
Page 100: ...Appendix F Technical Drawings F 22...
Page 101: ...Appendix F Technical Drawings F 23...
Page 102: ...Appendix F Technical Drawings F 24...
Page 103: ...Appendix F Technical Drawings F 25...
Page 104: ...Appendix F Technical Drawings F 26...
Page 105: ...Appendix F Technical Drawings F 27...
Page 106: ...Appendix F Technical Drawings F 28...
Page 107: ...Appendix F Technical Drawings F 29...
Page 108: ...Appendix F Technical Drawings F 30...
Page 109: ...Appendix F Technical Drawings F 31...
Page 110: ...Appendix F Technical Drawings F 32...
Page 116: ...Appendix G Calculation of Maunal Calibration Parameters G 6...