Appendix H. Error Message
H-2
3. Warning: Add Or Subtract
More Material . .
Explanation
While performing an Auto Calibration, you
added or removed material from the vessel.
The amount of material you moved produced
a change of less than 9000 counts. The
Weigh II will go ahead and calibrate the
system with the entered values. However, for
good calibration accuracy, a larger change in
counts/larger movement of material is
recommended.
Solution
Verify that the correct amount of material was
actually moved. Recalibrate when possible
with a larger movement of material.
4. All Channels are Hidden
Explanation
Vessel monitoring results are not shown on
the display because all channels are hidden.
Solution
Display all hidden channels again by
disabling
Hide
.
1. Put the Weigh II in Manual Mode.
2. Press the Shift Key (Shift LED illuminated)
and then press the ‘9’ Key.
3. The display acknowledges that
Hide
is
disabled. You can now scroll through all of
the channels in Manual Mode or switch to
Auto Mode.
5. All Channels are Disabled
Explanation
The Weigh II is not monitoring any channels
because all channels are disabled.
Solution
Display selected disabled channels again by
turning
Enab
off.
1. Put the Weigh II in Manual Mode.
2. Follow the path
Main/Service/ADC/Enab
.
3. Use the Up and Down Arrow Keys to scroll
to the channel that you want to enable.
Press the F1 Key to turn the channel on.
4. Repeat Step 3 as desired for other chan-
nels.
6. ADC Overrange (+/-)
Explanation
The sensor input is causing the counts to go
above 2,097,151 (if +) or to 0 (if -) because
one of the legs of the excitation voltage is not
functioning.
Solutions
1. Check the wiring of the sensor to
the junction box and junction box to
the Weigh II. Correct loose, damaged
(shorted), or incorrect wiring.
2. If there is no problem with the wiring, one
or more sensors may be damaged. Follow
troubleshooting procedures detailed in the
sensor manual to
locate the damaged sensor(s). Replace
if required.
3. If there is no problem with the sensors or
the wiring, check the power supply by
measuring across the +Ex and -Ex on the
Weigh II.
• If the Weigh II is wired to Kistler-Morse
sensors/transducers, verify that the
voltage is approximately 12 volts.
• If the Weigh II is wired to foil-gage
transducers, verify that the voltage
meets the transducer manufacturer’s
requirements.
If the voltage is incorrect, adjust
AdjEx
(
Main/Service/ADC/AdjEx
) until the correct
voltage is measured. If you cannot obtain
the correct voltage, replace the power
supply.
4. If Steps 1 through 3 indicated no problem,
measure the millivolts across the In+ and
In- terminals on the motherboard. This
reading is the actual signal coming back
from the sensor/transducer. For Kistler-
Morse sensors/transducers, the reading
should be less than 1 volt.
• If the reading is less than 1 volt, check
Gain
(
Main/Service/ADC/Gain
). If
Gain
is set correctly, the A/D converter is
probably damaged. Consult with
Kistler-Morse.
• If the voltage is greater than 1 volt,
recheck wiring, sensors, and power
supply as described in Steps 1
through 3.
Summary of Contents for Weigh II
Page 12: ...2 4 Chapter 2 Hardware Installation...
Page 20: ...3 8 Chapter 3 Menu Tree Keyboard Functions and Quick Start...
Page 50: ...6 12 Chapter 6 Calibration...
Page 68: ...A 2 Appendix A Product Specifications...
Page 70: ...B 2 Appendix B Summary of Commands Used When Vessel Mounting Hot Keys...
Page 77: ...Appendix E E 1 This appendix not used at this time...
Page 78: ...Appendix E E 2...
Page 80: ...Appendix F Technical Drawings F 2...
Page 81: ...Appendix F Technical Drawings F 3...
Page 82: ...Appendix F Technical Drawings F 4...
Page 83: ...Appendix F Technical Drawings F 5...
Page 84: ...Appendix F Technical Drawings F 6...
Page 85: ...Appendix F Technical Drawings F 7...
Page 86: ...Appendix F Technical Drawings F 8...
Page 87: ...Appendix F Technical Drawings F 9...
Page 88: ...Appendix F Technical Drawings F 10...
Page 89: ...Appendix F Technical Drawings F 11...
Page 90: ...Appendix F Technical Drawings F 12...
Page 91: ...Appendix F Technical Drawings F 13...
Page 92: ...Appendix F Technical Drawings F 14...
Page 93: ...Appendix F Technical Drawings F 15...
Page 94: ...Appendix F Technical Drawings F 16...
Page 95: ...Appendix F Technical Drawings F 17...
Page 96: ...Appendix F Technical Drawings F 18...
Page 97: ...Appendix F Technical Drawings F 19...
Page 98: ...Appendix F Technical Drawings F 20...
Page 99: ...Appendix F Technical Drawings F 21...
Page 100: ...Appendix F Technical Drawings F 22...
Page 101: ...Appendix F Technical Drawings F 23...
Page 102: ...Appendix F Technical Drawings F 24...
Page 103: ...Appendix F Technical Drawings F 25...
Page 104: ...Appendix F Technical Drawings F 26...
Page 105: ...Appendix F Technical Drawings F 27...
Page 106: ...Appendix F Technical Drawings F 28...
Page 107: ...Appendix F Technical Drawings F 29...
Page 108: ...Appendix F Technical Drawings F 30...
Page 109: ...Appendix F Technical Drawings F 31...
Page 110: ...Appendix F Technical Drawings F 32...
Page 116: ...Appendix G Calculation of Maunal Calibration Parameters G 6...