42
TOS5200
Setting Test Conditions
Upper limit (UPPER)
Sets the upper current limit that is used in judgments. If a current that is greater than or equal
to the upper limit is measured, a U-FAIL judgment occurs.
Lower limit (LOWER)
Sets the lower current limit that is used in judgments. While LOWER is on, if a current that is
less than or equal to the lower limit is measured, an L-FAIL judgment occurs.
You can turn on and off the lower limit function using the ON/OFF (SHIFT+UPR/LWR) key.
Test time (TEST)
Set the test time. The test time begins when the voltage rise time elapses and ends when the
specified time elapses.
You can turn on and off the test time using the ON/OFF (SHIFT+TEST) key.
If the test time is set to off, PASS judgment is not performed.
Press STOP to stop testing.
Voltage rise time (RISE)
Separate from the test time, you can set the time that the TOS5200 takes to raise the voltage
to the test voltage.
Test conditions (additional items)
Press MORE to display the items that are not displayed.
Upper limit
0.01 mA to 110 mA
Lower limit
0.01 mA to 110 mA
Test time
0.1 s to 999 s
Voltage rise time
0.1 s to 10.0 s
MORE
A
kV
M
ȍ
mA
CONFIG
RISE TEST
RMS AVE
MORE
CAL RMT
MT
LIMIT
50Hz 60Hz
UPPER
LOWER W COMP
s
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
Start voltage setting [0: off, 1: on]
Frequency setting [0: 50 Hz, 1: 60 Hz]
Measurement mode setting [0: RMS, 1: AVE]
Voltage fall time setting: [0: off, 1: on]
1
MEMORY
2
3
RECALL
Example: Selecting the frequency (blinking)
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......