66
TOS5200
Specifications
Other features
TOS5200
Test mode
Double action feature
Tests can only be started by pressing and releasing STOP and then pressing START
within 0.5 seconds of releasing the STOP switch.
Length of time to hold a PASS
judgment result
You can set the length of time to hold a PASS judgment: 50 ms, 100 ms, 200 ms, 1 s, 2 s,
5 s, or HOLD.
Momentary feature
Tests are only executed while the START switch is held down.
Fail mode feature
This feature enables you to prevent remotely transmitted stop signals from clearing FAIL
judgments and PROTECTION modes.
Timer feature
The test ends when the specified time elapses.
Output voltage monitor feature
(Volt Error)
If output voltage exceeds “s 350 V” or is lower than “setting - 350 V,” the TOS5200
switches to PROTECTION mode, output is turned off, and testing finishes.
Memory
Up to three sets of test conditions can be saved to memory.
Key lock
Locks panel key operations (settings and changes).
Protection functions
Under any of the following conditions, the TOS5200 switches to the PROTECTION state,
immediately turns output off, and stops testing. PROTECTION on the screen lights.
Interlock Protection
An interlock signal has been detected.
Power Supply Protection
An error was detected in the power supply.
Volt Error Protection
While monitoring the output voltage, a voltage outside of the rated limits was detected.
Withstanding voltage test: ±350 V
Over Load Protection
While monitoring the output power, power exceeding the output power limit was detected.
Withstanding voltage test: 550 VA
Over Heat Protection
The internal temperature of the TOS5200 became too high.
Over Rating Protection
The output current was generated for a length of time that exceeds the regulated time.
Remote Protection
A connection to or disconnection from the front-panel REMOTE connector was detected.
SIGNAL I/O Protection
The rear-panel SIGNAL I/O connector’s ENABLE signal has changed.
USB Protection
The USB connector has been disconnected, or a defect was detected during remote con-
trol operation.
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......