26
TOS5200
Connecting to the Device under Test (DUT)
Using the optional high voltage test probe
(model HP01A-TOS/HP02A-TOS)
If you use the optional test probe instead of the test leads, you can use hands-on control to
start tests. For details, see the “OPERATION MANUAL HIGH VOLTAGE TEST PROBE
HP01A-TOS/HP02A-TOS.”
Disconnecting test leads from the DUT
1
Check that the POWER switch is off and that the DANGER LED is off.
2
Disconnect the high-voltage test lead (red) from the front-panel HIGH
VOLTAGE terminal.
3
Disconnect the high-voltage test lead (red) from the DUT.
4
Disconnect the low-voltage test lead (black).
You can disconnect the low-voltage test lead (black) first from either the DUT or the
TOS5200.