62
TOS5200
Pre-Test Inspection
Inspection of test leads and the judgment feature
Allow a current to flow between the HIGH VOLTAGE and LOW VOLTAGE terminals to check
the judgment feature. During this inspection, the two test leads are shorted, so you can check
for breaks in the test leads at the same time.
We recommend that you perform this pre-test inspection before you use the TOS5200.
The test leads are consumable parts. Check them periodically for tears and breaks.
1
Short the low-voltage test lead (black) and high-voltage test lead (red).
2
Connect the low-voltage test lead (black) to the LOW VOLTAGE terminal.
3
Connect the high-voltage test lead (red) to the HIGH VOLTAGE terminal.
4
Set the test conditions.
Be sure to set the upper limit (UPPER).
5
Press START to start a test.
6
If a test results in a U-FAIL judgment, based on this simple check, both
the test leads and the TOS5200 judgment feature are working properly.
If a U-FAIL result does not occur even after you perform this check multiple times, there may
be breaks in the test leads. Double-check whether this is the case. If there are no breaks in
the test leads, the TOS5200 needs to be repaired.
To have your TOS5200 repaired, contact your Kikusui agent or distributor.
Calibration
The product is calibrated before shipment. To maintain long-term reliable performance, we
recommend periodic calibration.
To have your product calibrated, contact your Kikusui distributor or agent.
WARNING
Breaks or tears in the insulation may cause electric shock or fire. If a break or tear is
found, stop using the damaged test lead immediately.
See
See
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......