SIGNAL I/O Connector
6
External Co
ntrol
■
SIGNAL I/O connector pinout
Pin no.
Signal name
I/O
TOS5200
1
IN
I
If you open the positive and negative terminals, the output is turned off, and the TOS5200 is
switched to Protection mode.
Open: The resistance between the two terminals is 1.2 k
Ω
or greater.
Short: The resistance between the two terminals is 1 k
Ω
or less.
2
PM0
I
Panel memory selection signal.
The selection signal is latched on the ris-
ing edge of the input strobe signal to
recall panel memory.
* The selection of memory is prioritized
over TEST SEL and AUTO SEL.
3
PM1
I
4
NC
—
—
5
NC
—
—
6
NC
—
—
7
NC
—
—
8
NC
—
—
9
STB
I
Panel memory strobe signal input terminal.
10
TEST SEL
I
NA
11
AUTO SEL
I
NA
12
COM
—
Circuit common terminal.
13
INTERLOCK-
I
If you open the positive and negative terminals, the output is turned off, and the TOS5200 is
switched to Protection mode.
Open: The resistance between the two terminals is 1.2 k
Ω
or greater.
Short: The resistance between the two terminals is 1 k
Ω
or less.
14
HV.ON
O
On during testing and when a voltage remains across the output terminals.
15
TEST
O
On during testing (excluding when voltage is rising or falling).
16
PASS
O
On for at least 0.2 seconds (the PASS HOLD time) when a PASS judgment occurs.
On continuously when the PASS HOLD time is set to HOLD.
17
U-FAIL
O
On continuously when an UPPER FAIL judgment occurs because a value greater than or
equal to the upper limit is detected.
18
L-FAIL
O
On continuously when LOWER FAIL results from judgment because a value less than or
equal to the lower limit is detected.
19
READY
O
On when the TOS5200 is waiting (when it is in the READY state).
20
PROTECTION
O
On when protective features have been activated.
21
START
I
Start signal input terminal.
22
STOP
I
Stop signal input terminal.
23
ENABLE
I
Input terminal for the start signal’s ENABLE signal. If the ENABLE signal changes, the
TOS5200 is switched to Protection mode.
24
+24 V
—
+24 V internal power supply output terminal; maximum output current 100 mA.
25
COM
—
Circuit common terminal.
PM0
PM1
Recalled panel memory number
H
H
Memory 1
L
H
Memory 2
H
L
Memory 3
L
L
—
NC
NC
NC
NC
PM1
PM0
IN
NC
STB
TEST SEL
AITO SEL
COM
INTERLOCK-
HV.ON
TEST
PASS
U-FAIL
L-FAIL
READY
COM
+24V
ENABLE
STOP
START
PROTECTION
13 12 11 10
19 18 17 16 15 14
1
2
3
4
5
6
7
8
9
24 23 22 21 20
25
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......