46
TOS5200
CONFIG Settings
Buzzer volume level for FAIL judgment (FAIL)
You can set the volume level of the buzzer that is sounded when a FAIL judgment occurs.
RS232C baudrate
You can set the baud rate for the RS232C interface.
Talk mode
This feature enables you to send test results and the like from the product through the
RS232C interface without using commands from a PC. It can reduce processing on the PC.
If you turn talk mode on, you cannot control the product from a PC. To control the product
from a PC, turn talk mode off.
CONFIG 2 settings
On the CONFIG
setup screen, you can set the output conditions when using the optional
warning light unit, PL02-TOS.
H.V ON
Sets whether the 24 Vdc output is generated while there is a residual voltage and during test-
ing.
0 to 9
0, 1, 2, 3, 4, 5 (factory default setting), 6, 7, 8, 9
9600 (0) to 115200 (4)
0: 9600 (factory default setting), 1: 19200 , 2: 38400, 3: 57600,
4: 115200
See
off (0)
Talk mode is not used (factory default setting).
on (1)
Talk mode is used.
2
A
kV
M
ȍ
mA
CONFIG
RISE TEST
RMS AVE
MORE CAL RMT
MT
LIMIT
50Hz 60Hz
UPPER
LOWER W COMP
s
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
Ready setting [0: off, 1: on]
Lower Fail setting [0: off, 1: on]
Power ON setting [0: off, 1: on]
Protection setting [0: off, 1: on]
Pass setting [0: off, 1: on]
Test setting [0: off, 1: on]
H.V ON setting [0: off, 1: on]
Upper Fail setting [0: off, 1: on]
1
MEMORY
2
3
RECALL
Example: Selecting H.V ON (blinking)
off (0)
The 24 Vdc output is turned off while there is a residual voltage and during testing
(factory default setting).
on (1)
The 24 Vdc output is turned on while there is a residual voltage and during testing.
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......