Troubleshooting
App
PROTECTION mode cannot be cleared
Unable to use remote control
Symptom
Items to check and possible remedy
PROTECTION mode cannot be
cleared even if I press STOP and
restart the TOS5200.
Is an interlock signal being applied?
Stop the interlock signal input.
Is the internal temperature too high?
Confirm that the internal temperature of the TOS5200 has decreased, and then
clear the mode.
—
Is the REMOTE connector connected properly?
Check the REMOTE connector.
During remote control operation, did the USB or RS232C connector become
disconnected or are there problems with the connector?
Check the USB or RS232C connector.
Symptom
Items to check and possible remedy
Remote control is not possible
through the RS232C interface.
Is the baudrate set correctly?
Check the value.
See
See