TOS5200
67
Specifications
8
S
p
ecif
ica
tions
Interfaces
TOS5200
USB
USB Specification 2.0
Standard type B socket
RS232C
1
Hardware
D-SUB 9-pin connector (EIA-232-D compliant)
All functions except the POWER switch and key lock
Baudrate: 9600, 19200, 38400, 57600, 115200 bps
Transmission system: Start-stop synchronization
Data length: 8 bits, Stop bits: 1 bit, Parity bit: None
Program message
terminator
CR+LF during transmission, CR, LF, or CR+LF during reception
REMOTE
Front-panel 9-pin MINI DIN connector.
By connecting an optional device to this connector, you can control the starting and
stopping of tests remotely.
SIGNAL I/O
Rear-panel D-sub 25-pin connector
Output
specifications
Output method
Open collector output (4.5 Vdc to 30 Vdc)
Output withstanding
voltage
30 Vdc
Output saturation volt-
age
Approx. 1.1 V (25 °C)
Maximum output current
400 mA (TOTAL)
Input
specifications
2
High-level input voltage
11 V to 15 V
Low-level input voltage
0 V to 4 V
Low-level input current
5 mA max.
Input time width
5 ms minimum
1 IN
I
If you open the positive and negative terminals, the output is turned off, and the
TOS5200 is switched to Protection mode.
Open: If the resistance between the terminals is 1.2 k
Ω
or greater.
Short: If the resistance between the terminals is 1 k
Ω
or less.
2 PM0
Panel memory selection signal.
The selection signal is latched
on the rising edge of the input
strobe signal to recall panel
memory.
* The selection of memory is pri-
oritized over TEST SEL and
AUTO SEL.
I
3 PM1
I
4 NC
—
5 NC
—
6 NC
—
7 NC
—
8 NC
—
9 STB
I
Panel memory strobe signal input terminal.
10 TEST SEL
I
NA
11 AUTO SEL
I
NA
12 COM
—
Circuit common terminal.
13 INTERLOCK-
I
If you open the positive and negative terminals, the output is turned off, and the
TOS5200 is switched to Protection mode.
Open: If the resistance between the terminals is 1.2 k
Ω
or greater.
Short: If the resistance between the terminals is 1 k
Ω
or less.
14 HV.ON
O
On during testing and when a voltage remains across
the output terminals.
15 TEST
O
On during testing (excluding when voltage is rising or falling).
16 PASS
O
On for approximately 0.2 seconds when a PASS judgment occurs.
On continuously when the PASS HOLD time is set to HOLD.
17 U-FAIL
O
On continuously when UPPER FAIL results from judgment because a value greater
than or equal to the upper limit is detected.
PM0
PM1
Recalled panel memory number
H
H
Memory 1
L
H
Memory 2
H
L
Memory 3
L
L
—
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......