TOS5200
41
5
Withst
andin
g
V
o
lt
age
T
e
st
Setting Test Conditions
For details on how to select settings and enter values, see “Panel Operation Basics”.
For details on how to connect the TOS5200 to the DUT, see “Connecting to the Device under
Test (DUT)”.
For details on the LOWER and UPPER settings, see also “About Judgment”.
Test conditions (basic items)
Test voltage (Voltage)
Sets the test voltage. You cannot specify a voltage that is greater than or equal to the limit
voltage.
Limit voltage (LIMIT)
Sets the limit voltage. This prevents an unnecessarily high voltage from being applied to the
DUT because of incorrect operations of the TOS5200.
Test condition
Panel operation
Basic item
Test voltage (Voltage)
SET key, rotary knob
Limit voltage (LIMIT)
LIMIT (SHIFT+SET) key, rotary knob
Upper limit (UPPER)
UPR/ LWR key, rotary knob
Lower limit (LOWER)
UPR/ LWR key, rotary knob
on/ off
ON/ OFF (SHIFT+UPR/ LWR) key
Test time (TEST)
TEST key, rotary knob
on/ off
ON/ OFF (SHIFT+ TEST) key
Voltage rise time (RISE)
TEST key, rotary knob
Additional items
Test voltage frequency (Frequency)
MORE key, rotary knob
Start voltage (Start Voltage)
MORE key, rotary knob
Voltage fall time (Fall Time)
MORE key, rotary knob
Voltage/current measurement mode
(Measurement)
MORE key, rotary knob
See
See
See
A
kV
M
ȍ
mA
CONFIG
RISE TEST
RMS
AVE
MORE CAL RMT
MT
LIMIT
50Hz
60Hz
UPPER
LOWER W COMP
s
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
SET
UPR/LWR
TEST
MORE
LIMIT
ON/OFF
ON/OFF
KEY LOCK
LOCAL
SHIFT
Test voltage
0.00 kV to 5.50 kV
Limit voltage
0.00 kV to 5.50 kV
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......