TOS3200
57
Setting Test Conditions of the PCC Measurement
4
TC and
PCC T
est
s
Blinking indication UP <= LOW
If the lower reference is set to a value greater than or equal to the upper reference in the
LOWER ON condition, “UP <= LOW” will blink at the upper right of the screen to indicate that
the setting is invalid.
Setup Items of PCC2/2 (MODE / RANGE)
Press the NEXT (SHIFT+ ) key to display the PCC2/2 screen.
Table 4-6
Determination of the measurement range
SHIFT
PREV
NEXT
PCC 2 / 2
RMS
AUTO
MODE RANGE
MODE
RANGE
Item
Description
Panel operation
Selects the current measurement mode.
MODE (F2) key
RMS
RMS measurement
MODE
(SHIFT+F2) key
DC
DC measurement
PEAK
Peak measurement
Selects the measurement range.
RANGE (F3)
key
AUTO
Automatically switches the range according to the measured
value
RANGE
(SHIFT+F3) key
FIX
Fixes the range.
The measurement range is determined by the upper reference
and current measurement mode (MODE) settings. See Table 4-
6.
MODE
RANGE
Measurement
range
MODE
Upper reference
Range 1
DC, RMS
30 µA to 600 µA
PEAK
50 µA to 850 µA
Range 2
DC, RMS
601 µA to 6.00 mA
PEAK
851 µA to 8.50 mA
Range 3
DC, RMS
6.01 mA to 30.0 mA
PEAK
8.51 mA to 90.0 mA
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......