64
TOS3200
Stopping the Test
Aborting the Test
Press the STOP switch.
When the Test Ends
The test will stop in any of the conditions below.
a. When the test time has elapsed (when TIMER is ON).
b. When a current greater than or equal to the upper reference is measured (U-FAIL).
c. When you press the STOP switch.
When the test stops, the DANGER lamp will turn off, and the power line to the EUT will be
turned off.
If the test stops in the case of condition a or b above, the judgment result will be displayed on
the screen.
For details on how the TOS3200 judges the measured values, see “Judgment System.”
Table 4-7 shows the TOS3200 operation when the test ends for each judgement result.
Table 4-7
Operation when the test ends
*1
The duration of the buzzer for a PASS judgment is fixed to 0.2 s. It does not depend on the
PASS HOLD time.
Storing the Test Result
The storage function is assigned to a function key while the judgment result is displayed. For
details, see “Storing the Test Results.”
Releasing the Judgment Result
Press the STOP switch to set the TOS3200 to READY status.
START
STOP
TC
PASS
E
NC
P
E
OFF
ON
PROBE LOWER
NORM
POL
NORM
LINE
0.0W
0.0V 0.00A
NORM
COND
UPPER
ON
TIMER
0
µA
10
s
SAVE?
Yes
No
See
Operation
PASS
L-FAIL
U-FAIL
Display
Displays “PASS” on the
screen.
Displays “
↓
FAIL” on the
screen.
Displays “
↑
FAIL” on the
screen.
Buzzer:
Sounds for 0.2 s.
*1
Sounds until FAIL is cleared.
SIGNAL I/O
connector
Outputs the pass signal for
the time specified by PASS
HOLD.
Outputs the L-FAIL signal
until FAIL is cleared.
Outputs the U-FAIL signal
until FAIL is cleared.
See
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......