78
TOS3200
Creating Sequence Programs
Setup Items of AUTO1/2
AUTO 1 / 2
NTWK:A MODE:RMS ABORT:OFF
00 TC•E
NC
P
E
––P
NRM
•NORM––– 1s
01
TC•E
NC
P
E
––P
NRM
•NORM––– 1s
–––END
PRG
00:--UNTITLED--
EDIT
INS
UPPER
LOWER
WAIT
TIMER
LOWER : 30μA
UPPER : 30.0mA
WAIT
: OFF
TIMER : 1s
The number of the selected step is highlighted.
Program edit status
Test conditions of the
selected step
1 2
3 4
5 6
Test conditions common to all steps
(Set in AUTO2/2.)
7
8
9
10
You can move to each item using
the arrow keys (
).
Item
Description
Panel operation
1
(Step number) Selects the step number to be edited.
Rotary knob
2
(Operation
mode)
Selects TC or PCC.
Rotary knob
3
(
PROBE
)
*1
Selects the connection destination of measurement terminals A and B.
E
NC
P
E
Connected between the enclosure and earth.
Rotary knob
E
NC
E
NC
Connected between two enclosures.
E
NC
L
IV
Connected between the enclosure and power line (live).
E
NC
N
EU
Connected between the enclosure and power line (neu-
tral).
4
(
POL
)
*2
Selects the polarity of the power line supplied to the EUT.
P
NRM
Normal phase connection
Rotary knob
P
RVS
Reverse phase connection
5
(
COND
)
*2
Selects the single fault mode.
NORM
Normal status
Rotary knob
FLT
NEU
Power line (neutral) disconnected status
FLT
PE
Earth line disconnected status
6
(Interval)
Sets the time until the next step is executed in the range of 1 s to 99 s
or HOLD. If set to HOLD, the next step will not start until you press the
START switch.
Rotary knob
7
LOWER
Sets the lower limit of the judgment reference.
F2 key
Lower ref-
erence
The selectable range varies depending on the NTWK
(measurement network) and MODE (current measure-
ment mode) settings on AUTO2/2. See Table 6-1.
You will not be able to enter the value if set to OFF.
Rotary knob
ON/OFF
The item is turned on when the value is displayed.
SHIFT+F2 key
8
UPPER
Sets the upper limit of the judgment reference.
F3 key
Upper ref-
erence
The selectable range varies depending on the NTWK
(measurement network) and MODE (current measurement
mode) settings on AUTO2/2. See Table 6-1.
You will not be able to enter the value if set to OFF.
Rotary knob
ON/OFF
The item is turned on when the value is displayed.
SHIFT+F3 key
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......