96
TOS3200
SIGNAL I/O Connector
Internal Construction
The common for the input signal circuit and output signal circuit is the same.
It is pulled up to +12 V. If the input terminal is opened, the input signal circuit will be in the
same condition as when a high level signal is applied.
Fig. 7-1
Internal construction of SIGNAL I/O
Application Examples of Input Signal
Controlling the TOS3200 using a make contact
The input terminal is set to low level by using a make contact such as
a relay or switch.
Controlling the TOS3200 using a logical device
In place of a switch in the example above, a logical device such as
a transistor is used.
Construct the circuit so that at least 5 mA of collector current ic of
the transistor can flow.
Output
signal
controller
Input
signal
controller
COM
Output signal 1
Output signal 2
Input signal 1
Input signal 2
+15 V
+15 V
COM
S1
COM
ic
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......