TOS3200
97
Starting the Test
7
External Contr
o
l
Application Examples of Output Signal
Driving a relay
The output signal is used to drive a relay.
Producing a low level digital signal
The output signal is used to produce a low
level digital signal.
Starting the Test
To start the test using the SIGNAL I/O connector, set the ENABLE signal to low level. When
at least 10 ms passes after the READY signal is set to low level, set the START signal to low
level for at least 5 ms. When a valid START signal is detected, the READY signal will be set
to high level.
If the ENABLE signal is low level, the start signal of the SIGNAL I/O connector and the
START input of the REMOTE terminal will be enabled, and the START switch on the panel
will be disabled.
Fig. 7-2
START signal
COM
30 V or less
400 mA or less
Relay
COM
30 V or less
R
A
Low when outputting
a signal
READY (OUT)
ENABLE (IN)
START (IN)
10 ms or more
5 ms or more