52
TOS3200
Setting Test Conditions of the TC Measurement
Setup Items of TC1/2 (LOWER / UPPER / TIMER / WAIT)
For a description of the LOWER, UPPER, and TIME/WAIT settings, see also “Judgment Sys-
tem.”
Table 4-3
Selectable range of LOWER/UPPER
Blinking UP <= LOW indication
If the lower reference is set to a value greater than or equal to the upper reference in the
LOWER ON condition, “UP <= LOW” will blink at the upper right of the screen to indicate that
the setting is invalid.
See
TC 1 / 2
READY
E
NC
P
E
OFF
ON
PROBE LOWER
UPPER TIMER
POL
NORM
NORM
COND
UPPER
OFF
TIMER
30.0
mA
10
s
PROBE
LOWER
UPPER
TIME/WAIT
Item
Description
Panel operation
Sets the lower limit of the judgment reference. If LOWER ON is specified and a current
less than or equal to the lower reference is measured, L-FAIL judgment will result
LOWER (F3) key
Lower
reference
The selectable range varies depending on the NTWK
(measurement network) and MODE (current measurement
mode) settings on TC2/2. See Table 4-3.
Rotary knob
OFF
Disables the lower reference judgment.
LOWER ON/OFF
(SHIFT+F3) key
ON
Enables the lower reference judgment.
Sets the upper limit of the judgment reference. If UPPER ON is specified and a current
greater than or equal to the upper reference is measured, U-FAIL judgment will result.
UPPER (F4) key
Upper
reference
The selectable range varies depending on the NTWK
(measurement network) and MODE (current measurement
mode) settings on TC2/2. See Table 4-3.
Rotary knob
OFF
Disables the upper reference judgment.
UPPER ON/OFF
(SHIFT+F4) key
ON
Enables the upper reference judgment.
Sets the test time (TIMER) or the test wait time (WAIT). If TIMER is selected, you can
set the test time and TIMER ON/OFF. If WAIT is selected, you can set the test wait time
and WAIT ON/OFF.
TIME/WAIT (F5)
key
Test time
Set the test time in the range of 1 s to 999 s.
The test time will start when the test wait time elapses.
Rotary knob
TIMER OFF
The specified test time is discarded.
Press the STOP switch to stop the test.
TIMER ON/OFF
(SHIFT+F5) key
TIMER ON
The test will stop when the specified time elapses.
Test wait
time
Set the time until the test is actually started after the START
switch is pressed (test wait time) in the range of 1 s to 999 s.
Rotary knob
WAIT OFF
The specified test wait time is discarded.
WAIT ON/OFF
(SHIFT+F5) key
WAIT ON
The actual test will start when the specified test wait time
elapses.
LOWER
UPPER
TIMER
WAIT
MODE
NTWK A, B, B1,C
NTWK D, E
NTWK F
NTWK G
DC, RMS
30 µA to 30.0 mA
30 µA to 30.0 mA
30 µA to 20.0 mA
30 µA to 15.0 mA
PEAK
50 µA to 90.0 mA
50 µA to 45.0 mA
50 µA to 30.0 mA
50 µA to 22.5 mA
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......