Keysight PD1000A Power Device Measurement System Startup Guide
71
Running the S-Parameter Characterization
On State and Off State S-Parameter Characterization
4
Configure Hardware
. Use the Configure Hardware dialog to specify test
instrument IP and VISA addresses, timeout, SMU channels, etc.
E5080A ENA Configuration
a
ENA VISA Address:
b
ENA VISA Timeout (in seconds):
Maximum time that the PD1000A waits for
the ENA to respond to commands. (Typical value is 30 seconds.)
B2902A Configuration (Only for On-State S-Parameter)
a
B2902A VISA Address:
Locate the B2902A VISA address in the Connection
Expert utility in the Keysight IO Libraries Suite.
b
B2902A VISA Timeout (in seconds):
Maximum time that the PD1000A waits for
the B2902A to respond to commands. 10 seconds is typical.
c
Drain SMU:
S Use the drop-down menu to choose which channel to use for
biasing the DUT drain/collector. Typically, this is
Channel 2
of the SMU.
d
Gate SMU:
Use the drop-down menu to choose which channel to use for
biasing the DUT Gate. Typically, this is
Channel 1
of the SMU.
e
Room temperature:
Specifies the temperature in °C for the room in which the
tests are made. (This information is used in the output .mdm file.)
f
To save settings and close the dialog, click the
X
icon in the title bar.
Figure 36
S-Parameter Hardware Configuration Screen
Summary of Contents for PD1000A
Page 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Page 2: ......
Page 4: ......
Page 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Page 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Page 107: ......