Keysight E1412A Digital Multimeter User Guide
xcix
CALibration
Multimeter Command Reference
CALibration
The
CALibration
command subsystem allows you to enter a security code to
prevent accidental or unauthorized calibrations of the multimeter. When you first
receive your multimeter, it is secured. You must unsecure it by entering the
correct security code before you can calibrate the multimeter (see
CALibration:SECure:STATe
command).
Subsystem Syntax
CALibration
:COUNt?
:LFRequency 50|60|400
:LFRequency? [MIN|MAX]
:SECure:CODE <
new code
>
:SECure:STATe OFF|ON,<
code
>
:SECure:STATe?
:STRing <
quoted string
>
:STRing?
:VALue <
cal_value
>
:VALue?
:ZERO:AUTO ON|OFF
:ZERO:AUTO?
:COUNt?
CALibration:COUNt?
queries the multimeter to determine the number of times a
point calibration has occurred. A complete calibration of the multimeter
increases the count by the number of points calibrated. It is not a record of
complete calibrations. The count is stored in non-volatile memory.
Comments
-
*RST
does not change the calibration count stored in non-volatile memory.
Example
Query the Number of Occurrences of Point Calibrations
CAL:COUN?
Query the calibration count.