Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
CURRent[:DC]:APERture?
[SENSe:]CURRent[:DC]:APERture? [MIN|MAX]
returns one of the following
numbers to the output buffer.
–
The present aperture time in seconds if
MIN
or
MAX
is not specified.
–
The minimum aperture time available (.333 ms) if
MIN
is specified.
–
The maximum aperture time available (1.67 s @ 60Hz; 2 s @ 50Hz) if
MAX
is specified.
Example
Query the Aperture Time
CURR:APER 167E-03
Aperture time is 167ms.
CURR:APER?
Query multimeter to return aperture time.
enter statement
Enter value into computer.
CURRent[:DC]:NPLC
[SENSe:]CURRent[:DC]:NPLCycles <
number
>
sets the integration time in power
line cycles (PLCs). Values are rounded up to the nearest number of PLCs shown
in the following table.
Parameters
Comments
–
MINimum
selects 0.02 PLCs.
MAXimum
selects 100 PLCs. Setting the
integration time in power line cycles (PLCs) also sets the aperture time and
the resolution. For example, 10 PLCs sets an aperture time of 167ms (60Hz
line frequency) or 200ms (50Hz). The corresponding resolution depends on
the function and range you select.
–
The
CURR:DC:NPLC
command overrides the results of previously executed
CURRent:APERture
or
CURRent:RESolution
command.
–
The greater the number of PLCs, the greater the normal mode rejection
(and the lower the reading rate).
–
Only the 1 PLC, 10 PLC and 100 PLC settings provide normal mode
rejection of 50Hz or 60Hz power line related noise. Fractional PLC settings
of 0.02 and 0.2 do not provide normal mode rejection of power line noise.
Parameter Name Parameter Type
Range of Values
Default
Units
<
number
>
numeric
0.02|0.2|1|10|100|MIN|MAX
PLCs