Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
OUTPut
The VXIbus trigger lines are open-collector TTL lines that remain in a
non-asserted (high) state until the
voltmeter complete
signal is sent.
More than one TTL output trigger line can be enabled at one time.
*RST Condition:
OUTP:TTLT
n
OFF
Example
Route Voltmeter Complete to Trigger Line
OUTP:TTLT7 ON
Route signal to trigger line 7.
:TTLTrg[:STATe]?
OUTPut:TTLTrg<
n
>[:STATe]?
returns a number to show whether VXIbus trigger
line routing of the
voltmeter complete
signal is enabled or disabled: “
1
” =
ON
,
“
0
” =
OFF
. The number is sent to the output buffer.
Example
Query Voltmeter Complete Destination
OUTP:TTLT7 ON
Route signal to trigger line 7.
OUTP:TTLT7?
Query multimeter to return trigger line mode.
enter statement
Enter value into computer.
20ms (50Hz)
20.5ms
370
s
16.7ms
(60Hz)
17.2ms
390
s
2.5ms
(400Hz)
3.1ms
430
s
100
s
520
s
250
s
10
s
70
s
Aperture
Time
Voltmeter Complete Low
Autozero ON
Autozero OFF