ccxxvi
Keysight E1412A Digital Multoimeter User Guide
Keysight E1412A Multimeter Error Messages
Execution Errors
You turned math on (
CALC:STAT ON
) and then changed to a math operation that
was not valid with the present measurement function. For example, dB
measurements are not allowed with 2-wire ohms. The math state is turned off as
a result of this condition.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example:
TRIG:COUN -3
-223
Too much data
A character string was received but could not be executed because the string
length was more than 12 characters. This error can be generated by the
CALibration:STRing
and
DISPlay:TEXT
commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the
command. You may have used an invalid parameter choice.
Examples:
CALC:FUNC SCALE
(
SCALE
is not a valid choice) or
SAMP:COUN
ON
(
ON
is not a valid choice).
-230
Data stale
A
FETCh?
command was received but internal reading memory was empty. The
reading retrieved may be invalid or settings have changed since the data was
taken.
-330
Self-test failed
The multimeter's complete self-test failed from the remote interface (
*TST?
command). In addition to this error, more specific self-test errors are also
reported.
See also “Self-Test Errors,” following this section.
-350
Too many errors
The error queue is full because more than 20 errors have occurred. No additional
errors are stored until you remove errors from the queue. The error queue is
cleared when power has been off, or after a
*CLS
(clear status) command has
been executed.
-410
Query
INTERRUPTED
A command was received which sends data to the output buffer, but the output
buffer contained data from a previous command (the previous data is not
overwritten). The output buffer is cleared when power has been off, or after a
*RST
(reset) command has been executed.