cxxx
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
MEASure
Parameters
Comments
–
To select a standard measurement range, specify
range
as the input
signal's maximum expected resistance. The multimeter then selects the
correct range that will accept the input.
–
The
AUTO
or
DEFault
option for the
range
parameter enables autorange.
The
DEF
option for the
resolution
parameter defaults the integration time
to 10 PLC.
–
The
MIN
and
MAX
parameters select the minimum or maximum values for
range
and
resolution
:
For
range
:
MIN
= 100
;
MAX
=100M
For
resolution
:
MIN
selects the best resolution (the smallest value) for the
selected range.
MAX
selects the worst resolution (the largest value) for the
selected range.
–
To select autorange, specify
AUTO
or
DEF
for
range
or do not specify a
value for the range and resolution parameters. In the autorange mode, the
multimeter samples the input signal before each measurement and selects
the appropriate range.
–
To specify a
MIN
or
MAX
resolution while autoranging, you must specify
MEAS:FRES? DEF
(you cannot omit the
range
parameter). This prevents
the
MIN
or
MAX
resolution from being interpreted as a range setting and
the resulting command becomes
MEAS:FRES? DEF,MIN
or
MEAS:FRES? DEF,MAX
.
–
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making 4-Wire Ohms Measurements
MEAS:FRES? 1500,MAX
Function: 4-wire ohms; range
selected: 10kW; MAX resolution: 1
Parameter Name
Parameter Type
Range of Values
Default
Units
<
range
>
numeric
100
|1k
|10k
|100k
|1M
|10M
|100M
|
MIN|MAX|DEF|AUTO
ohms
<
resolution
>
numeric
resolution
|MIN|MAX|DEF
ohms