cxxvi
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
MEASure
MEASure
The
MEASure
command subsystem configures the multimeter to perform the
specified measurement with the given range and resolution. When the
multimeter is triggered,
MEASure
makes the measurement and sends the
readings to the output buffer.
Executing
MEASure
is equivalent to configuring the multimeter with the low-level
commands shown in the following table.:
Subsystem Syntax
MEASure
:CURRent:AC? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:CURRent[:DC]? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:FREQuency? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:FRESistance? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:PERiod? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:RESistance? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:VOLTage:AC? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
[:VOLTage[:DC]]? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
[:VOLTage[:DC]]:RATio? [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|D
EF]]
Command
Setting
RANGe
As specified (or AUTO).
RESolution
As specified, or as a function of range, integration
time, or NPLCs.
AC filter
([SENSe:]DET:BAND)
20 Hz - 300 kHz (medium filter)
Autozero
([SENSe:]ZERO:AUTO)
OFF if resolution setting results in NPLC <1;
ON if resolution setting results in NPLC
1
Input resistance
([SENSe:]INP:IMP:AUTO)
Applies to dc voltage and is disabled for all other
functions. 10M
for all dc voltage ranges.
Samples per trigger
(SAMP:COUN)
Trigger count
(TRIG:COUN)
Trigger delay
(TRIG:DEL)
Trigger source
(TRIG:SOUR)
1 sample
1 trigger
AUTO (Automatic delay)
IMM (trigger signal is always true)
Math function
(CALCulate:STATe)
OFF
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter User Manual and SCPI Programming Guide 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 18: ...xviii Keysight E1412A Multimeter User Guide ...
Page 242: ......