cxiv
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
CONFigure
Parameters
Comments
-
To select a standard measurement range, specify
range
as the input signal's
maximum expected voltage. The multimeter then selects the correct range
that will accept the input.
-
The
AUTO
or
DEFault
option for the
range
parameter enables autorange. The
DEFault
option for
resolution
defaults the integration time to 10 PLC.
-
The
MIN
and
MAX
parameters select the minimum or maximum values for
range
:
For
range
:
MIN
= 0.1V;
MAX
= 300V.
For
resolution
: See Table 3-4 for valid resolution choices for each range.
-
To select autorange, specify
AUTO
or
DEF
for
range
or do not specify a value
for the range and resolution parameters. In the autorange mode, the
multimeter samples the input signal before each measurement and selects
the appropriate range.
-
To specify a
MIN
or
MAX
resolution while autoranging, you must specify
AUTO
or
DEFault
for range;
CONF:VOLT:AC DEF,MIN
or
CONF:VOLT:AC DEF,MAX
or
CONF:VOLT:AC AUTO,MIN
or
CONF:VOLT:AC AUTO,MAX
(you cannot omit
the
range
parameter). This prevents the
MIN
or
MAX
resolution from being
interpreted as a range setting.
Example
Making AC Voltage Measurements
CONF:VOLT:AC 0.54,MAX
Function: AC volts; range selected: 1A;
MAX resolution: 100
A.
SAMP:COUN 3
Take 3 readings; source is IMMediate by default.
READ?
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
enter statement
Enter readings into computer.
Parameter Name Parameter Type
Range of Values
Default
Units
<
range
>
numeric
0.1V|1V|10V|100V|300V|
MIN|MAX|DEF|AUTO
volts
<
resolution
>
numeric
resolution
|MIN|MAX|DEF
volts