Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
Parameters
Comments
–
To select a standard measurement range, specify range as the input
signal’s maximum expected resistance. The multimeter then selects the
correct range.
–
MIN
selects the minimum range available with the
RESistance:RANGe
command: 100
.
MAX
selects the maximum range available: 100M
.
–
You must select a range using
RESistance:RANGe
before specifying
resolution. Also, in order to specify an aperture time of 10ms, you must
select a fixed range.
–
Specifying a fixed range disables the autorange mode set by the
RES:RANG:AUTO
command.
–
The
RES:RANG
command overrides the range setting from a previous
CONFigure
command on the same function. The multimeter uses the same
aperture time to set the resolution on the new range as was selected by
CONFigure
.
–
*RST
Condition:
RES:RANG 1k
Example
Change the Range
CONF:RES 1320,MAX
Function: 2-wire ohms;
range selected: 10k
; MAX resolution: 1
.
RES:RANG 220
Range selected: 1k
; MAX resolution: 0.1
.
READ?
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
enter statement
Enter readings into computer.
RESistance:RANGe?
–
[SENSe:]RESistance:RANGe? [MIN|MAX]
returns one of the following
numbers to the output buffer:
–
The present resistance range selected if
MIN
or
MAX
is not specified. Only
the ranges available with the
RANGe
command are returned. For example,
if
CONFigure
selects the 900
range, 1k
is the range returned.
Parameter Name
Parameter Type
Range of Values
Default
Units
<
range
>
numeric
100
|1k
|10k
|100k
|1M
|10M
|100M
|MIN|MAX
ohms