Keysight E1412A Multimeter Error Messages
Calibration Errors
Calibration Errors
The following errors indicate failures that may occur during a calibration. The
most common errors have descriptions here. Refer to the
E1412A Service
Manual
for more information on the other errors.
701
Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the
multimeter. When applicable, this error will occur at power-on to warn you that
the multimeter is unsecured.
702
Cal secured
The multimeter is secured against calibration.
703
Invalid secure code
An invalid calibration security code was received when attempting to unsecure or
secure the multimeter. You must use the same security code to unsecure the
multimeter as was used to secure it, and vice versa. The security code may
contain up to 12 alphanumeric characters. The first character
must be
a letter.
704
Secure code too long
A security code was received which contained more than 12 characters.
705
Cal aborted
A calibration in progress is aborted when you send a device clear to the
multimeter.
706
Cal value out of range
The specified calibration value (
CAL:VALue
) is invalid for the present function and
range.
707
Cal signal measurement out of range
The specified calibration value (
CAL:VALue
) does not match the signal applied to
the multimeter.
708
Cal signal frequency out of range
The input signal frequency for an ac calibration does not match the required
input frequency for calibration.
709
No cal for this function or range
You cannot perform calibrations for ac current, period, continuity, diode, ratio, or
on the 100M
range.