Keysight E1412A Digital Multimeter User Guide
clxxxiii
[SENSe:]
Multimeter Command Reference
VOLTage[:DC]:RESolution?
[SENSe:]VOLTage[:DC]:RESolution? [MIN|MAX]
returns one of the following
numbers to the output buffer.
–
The present resolution selected if
MIN
or
MAX
is not specified. Only the
resolution values available on ranges set by the
RANGe
command are
returned.
–
The resolution with the smallest value (i.e., the best resolution) for the
selected range if
MIN
is specified.
–
The resolution with the largest value (i.e., the worst resolution) for the
selected range if
MAX
is specified.
Example
Query the Resolution
VOLT:DC:RES 1E-03
Set resolution to 1 mV.
VOLT:DC:RES?
Query multimeter to return the present resolution.
enter statement
Enter value into computer.
ZERO:AUTO
[SENSe:]ZERO:AUTO <
mode
>
enables or disables the autozero mode. Autozero
applies to dc voltage, dc current and 2-wire ohms measurements only. 4-wire
ohms and dc voltage ratio measurements automatically enable the autozero
mode.
Parameters
Comments
–
You can substitute decimal values for the
OFF
(“
0
”) and
ON
(“
1
”)
parameters.
–
The
ON
parameter enables autozero. This is the default parameter which
causes the multimeter to internally disconnect the input signal following
each measurement and make a zero measurement. The zero reading is
subtracted from the input signal reading to prevent offset voltages present
on the multimeter’s input circuitry from affecting measurement accuracy.
Parameter Name
Parameter Type
Range of Values
Default Units
<
mode
>
boolean
OFF|0|ON|1|ONC
E
none