Keysight E1412A Multimeter Application Information
Default Delays
If you do not specify a trigger delay, the multimeter automatically determines a
delay time (default delay) based on the present measurement function, range,
resolution, integration time and AC filter bandwidth setting. The delay time is
actually the settling time required before measurements to ensure measurement
accuracy. The default delay time is automatically updated whenever you change
the function or range. Once you specify a delay time value, however, the value
does not change until you specify another value, reset the multimeter or do a
CONF
or
MEAS
command. Table 8 shows the default trigger delay times for all
functions. This delay will occur before
each
measurement (see the trigger system
diagram in Figure 2-1).
You can specify a shorter delay time than the default values
shown. However, the shorter settling time may not produce
accurate measurements.