Keysight 41800A Active Probe
Adjustment
LF Gain/Flatness Adjustment
5-
8. Press the
TRIG/RESET
key, and the
'CONT mode'
softkey to select the
continuous sweep mode.
9. Adjust R18 (LF GAIN) and R25 (FLATNESS) so that the T/R data displayed
by the Keysight 4195A is within ±0.5 dB for the data at 50 MHz, as shown in
.
Figure 5-2
Adjustment Component Locations
Figure 5-3
Adjustment Example