40
Keysight 41800A Active Probe
Performance Tests
Probe Gain/Frequency Response Tests
4-
12.Sequentially press the following keys to set the Keysight 4195A's analysis
range.
MORE, 'ANA RNG'
(move the
*
marker to the 50 Hz point)
'active oMKR
*
MKR'
(
“oMKR”
in this softkey will change to
intensified green)
(move the o marker to the 10 Hz point)
'STORE ANA RNG'
'PART ANA on off'
(
“on”
in this softkey will change to
intensified green)
13.Press the
MKR
&
key and the
'MKR
&
MAX'
softkey to move the o marker to
the maximum point between 10 Hz and 50 Hz. And confirm that the T/R
data displayed on the upper right corner of the Keysight 4195A's display is
1 dB/–2 dB.
14.Press the
'MKR
&
MIN'
softkey to move the o marker to the minimum point
between 10 Hz and 50 Hz. And confirm that the T/R data displayed in the
upper right corner of the Keysight 4195A's display is 1 dB/–2 dB.
15.Sequentially press the following keys to change the Keysight 4195A's
analysis range.
MORE, 'ANA RNG'
(move the o marker to the 200 MHz point)
'STORE ANA RNG'
16.Press the
MKR
&
key and the
'MKR
&
MAX'
softkey to move the o marker to
the maximum point between 50 Hz and 200 MHz, and confirm that the T/R
data is within ±1 dB.
17.Press the
'MKR
&
MIN'
softkey to move the o marker to the minimum point
between 50 Hz and 200 MHz, and confirm that the data T/R is within ±1 dB.
18.Sequentially press the following keys to change the Keysight 4195A's
analysis range.
MORE, 'ANA RNG'
(move the o marker to the 200 MHz point)
'active oMKR
*
MKR'
(
“
*
MKR”
in this softkey will change to
intensified green)
(move the
*
marker to the 500 MHz point)
'STORE ANA RNG'