Keysight 41800A Active Probe
Operation Note
5 Adjustment
Introduction
This section describes the adjustments required to return the 41800A to the
specifications listed in
. If the 41800A failed the
performance test, or after it has been repaired. If proper performance cannot
be achieved after these adjustments, refer to
Chapter 6 , “Service”, on page
.
Equipment Required
lists the equipment required for adjustment.
Preparation
This paragraph provides the information which you should know, and the
procedures that you must perform before starting the adjustments.
1. The test equipment must be allowed to warm-up and stabilize for at least
30 minutes.
2. A Keysight 4195A is required to test the 41800A. In the remainder of this
manual, softkeys are indicated in boldface type and enclosed in single
quotes (e.g.,
'NETWORK'
softkey), and keys are indicated in boldface type
only (e.g.
PRESET
key).